2400 Series SourceMeter
®
User’s Manual Contact Check Function F-19
Use the following table to determine the parameter value for the
desired decimal digital output pattern. For non-decimal parame
-
ters, convert the decimal value to its binary, octal, or hexadecimal
equivalent.
Configure and control contact check event detection
SOURce <name>
:TRIGger:SEQuence2:SOURce <name> Enable/disable event detection
Parameters <name> = IMMediate Disable contact check event detection
CCHeck Enable contact check event detec-
tion
Query :SOURce? Query state of contact check event detec-
tion
Description This command enables or disables the contact check event detec-
tion. When event detection is enabled, the unit will wait the specified
timeout period (
“TOUT <NRf>,” page F-20) before issuing an open
lead failure and moving on to the next measurement.