User manual

F-8 Contact Check Function 2400 Series SourceMeter
®
User’s Manual
Figure F-4
Grading mode contact check limit testing
Perform
Limit 1 Test
?
Yes
No
Pass
?
No
Yes
Display
FAIL
Perform
Limit 2,3,5-12
Tests
?
Yes
No
Pass
?
No
Yes
Another
Test Cycle
?
No
Yes
Start
Turn Output ON and press LIMIT key.
Output First
Fail Pattern
Test
Another
Device
?
No
Yes
Stop
Press LIMIT key
Yes
No
Wait for SOT pulse
from handler
Perform Source-
Measure action
Display
FAIL
Binning
Control
Binning
Control
First
Failure
?
First
Failure
?
End
End
Yes
Yes
Output Limit 1
Fail Pattern
Immediate
Output Limit 2,3,5-12
Fail Pattern
Immediate
Store Limit 1 Fail
Pattern in Memory
Store Limit 2,3, 5-12 Fail
Pattern in Memory
No
No
Any
Failures
?
Display
PASS
Yes
Binning
Control
End
Any
Failures
?
Immediate
Output Pass
Pattern
No
Perform
Limit 4 Test
?
Yes
No
Pass
?
No
Yes
Store Limit 4 Fail
Pattern in Memory
No
First
Failure
?
Yes
Binning
Control
End
Immediate
Output Limit 4
Fail Pattern
Display
FAIL
Perform
Contact Check
NOTE Enabling contact check does not automatically enable LIMIT 4. LIMIT 4
must be explicitly enabled from the front panel or using the :CALC2 subsystem com-
mands. See“Front panel contact check operation,” page F-11, or “Remote contact
check operation,” page F-14 for more details.