User manual
A-8 Specifications 2400 Series SourceMeter
®
User’s Manual
Cases V and VI: Measure one function, Output Auto-Off disabled, and
no source setting changes
Figure A-5
Case V Timing Diagram
Auto-Zero: Enabled
Trigger Latency: 500µsec max
Figure A-6
Case VI Timing Diagram
Auto Zero: Disabled
Trigger Latency: 100µsec max
The source turns on as soon as the output is turned on and remains on until the
source is turned off. As shown in the two timing diagrams, the static source
remains on for every measurement cycle. The Source-Delay portion of the SDM
cycle is omitted. With Trigger Delay set to zero, Trigger Latency is the time from
when the trigger event occurs to when the SourceMeter begins an A/D conver-
sion.
Trigger
Event
Source On
Trigger
Delay
Trigger
Latency
A/D
Conversion
(ref zero phase)
A/D
Conversion
(ref phase)
A/D
Conversion
(signal phase)
Trigger
Event
Source On
Trigger
Delay
Trigger
Latency
A/D
Conversion
(signal phase)