User’s Manual SCPI Command Reference 18-113
Reset timestamp
RESet
:SYSTem:TIME:RESet Reset timestamp
Description This action command is used to reset the absolute timestamp to 0
seconds. The timestamp also resets to 0 seconds every time the
Source-Meter is turned on.
Auto reset timestamp
RESet:AUTO <b>
:SYSTem:TIME:RESet:AUTO <b> Reset timestamp when exiting idle
Parameters <b> = 1 or ON Enable auto timestamp reset
0 or OFF Disable auto timestamp reset
Query :AUTO? Query enabled/disabled auto timestamp
reset state
Description :RES:AUTO enables or disables auto timestamp reset. When
enabled, the timestamp will be automatically reset when exiting the
idle layer of the trigger model. This command is intended for use with
READ?/INIT when taking more than one reading.
Auto range change mode
RCMode <name>
:SYSTem:RCMode <name> Control auto range change mode
Parameters <name> = SINGle Single mode
MULTiple Multiple mode
Query :RCMode? Query auto range change mode
Description This command controls the auto range change mode. In the SINGle
mode, the SourceMeter will auto range only after first taking a read
-
ing. In the MULTiple mode, the SourceMeter will auto range up on
compliance in the Delay phase of the Source-Delay-Measure cycle,
thereby minimizing the chance of a SourceMeter being in compliance
in a multiple-SourceMeter system. A SourceMeter can downrange
only once a reading has been taken. In the MULTiple mode, you can
control the soak time using the :SOUR:SOAK command. (See
SOURce subsystem.”) Note that you can use the LLIMIT and
ULIMIT commands to control auto range limits. (See “SENSe1 sub-
system.”)