18-104 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
Description This command is used to select the power-on defaults. With RST
selected, the instrument powers up to the *RST default conditions.
With PRES selected, the instrument powers up to the :SYStem:PRE
-
Set default conditions. Default conditions are listed in the SCPI
tables (
Table 18-1 through Table 18-11).
With the SAV0-4 parameters specified, the instrument powers-on to
the setup that is saved in the specified location using the *SAV com
-
mand.
Control remote sensing
CCHeck <b>
:SYSTem:CCHeck <b> Enable or disable contact check.
Parameters <b> = 0 or OFF Disable contact check
1 or ON Enable contact check
Query CCHeck? Query state of contact check
Description This command is used to enable or disable contact check. When
using contact check, remote sensing must be enabled (:SYST:RSEN
ON).
RESistance <NRf>
:SYSTem:CCHeck:RESistance <NRf> Set contact check threshold resistance.
Parameters <NRf>= 0- 60 ohms (2 ohms, 15 ohms, or 50 ohms
threshold resistance)
0 to 2 = 2 ohms
2 to 15 = 15ohms
15 to 60 = 50 ohms
Query RESistance? Query contact check threshold resistance.
Description This command sets the contact threshold resistance value. (Factory
default = 50 ohms.) The contact check function will indicate a failure
when the connection path resistance is greater than the programmed
value. This command is used to enable or disable contact check.
When using contact check, remote sensing must be enabled
(:SYST:RSEN ON).
RSENse <b>
:SYSTem:RSENse <b> Enable or disable remote sensing.
Parameters <b> = 0 or OFF Disable remote sensing
1 or ON Enable remote sensing
Query :RSENse? Query state of remote sensing