User manual
18-102 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
Select default conditions
PRESet
:STATus:PRESet Return registers to default conditions
Description When this command is sent, the following SCPI event registers are
cleared to zero (0):
1. Operation Event Enable Register.
2. Event Enable Register.
3. Measurement Event Enable Register.
NOTE The Standard Event Register is not affected by this command.
Error queue
[:NEXT]?
:STATus:QUEue[:NEXT]? Read Error Queue
Description As error and status messages occur, they are placed into the Error
Queue. This query command is used to read those messages. See
Appendix B for a list of messages.
NOTE The :STATus:QUEue[:NEXT]? query command performs the same function
as the :SYSTem:ERRor? query command. (See
“SYSTem subsystem,”
page 18-103.)
CLEar
:STATus:QUEue:CLEar Clear Error Queue
Description This action command is used to clear the Error Queue of messages.
ENABle <list>
:STATus:QUEue:ENABle <list> Enable messages for Error Queue
Parameters <list> = (numlist)
where numlist is a specified list of messages that you wish to enable
for the Error.
Query :ENABle? Query list of enabled messages
Description On power-up, all error messages are enabled and will go into the
Error Queue as they occur. Status messages are not enabled and
will not go into the queue. This command is used to specify which
messages you want enabled. Messages not specified will be dis
-
abled and prevented from entering the queue.