18-94 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
SAVE <NRf>
:SOURce[1]:MEMory:SAVE <NRf> Save setup in specified memory location
Parameters <NRf> = 1 to 100 Specify memory location
Description This command is used to save the present instrument setup in a
memory location. Up to 100 setups can be saved. The following set
-
tings are saved in each source memory location:
SENSe[1]:CURRent:NPLCycles
SENSe[1]:RESistance:NPLCycles
SENSe[1]:VOLTage:NPLCycles
SENSe[1]:FUNCtion:CONCurrent
SENSe[1]:FUNCtion:ON
SENSe[1]:FUNCtion:OFF
SENSe[1]:RESistance:MODE
SENSe[1]:RESistance:OCOMpensated
SENSe[1]:AVERage:STATe
SENSe[1]:AVERage:TCONtrol
SENSe[1]:AVERage:COUNt
SOURce[1]:FUNCtion:SHAPe (Model 2430 only)
SOURce[1]:FUNCtion:MODE
SOURce[1]:DELay (Model 2430 in DC Mode)
SOURce[1]:DELay:AUTO (Model 2430 in DC Mode)
SOURce[1]...X...:TRIGgered:SFACtor
SOURce[1]...X...:TRIGgered:SFACtor:STATe
where: ...X... = :CURRent or :VOLTage (based on source mode)
SOURce[1]:PULSe:WIDTh (Model 2430 Pulse Mode only)
SOURce[1]:PULSe:DELay (Model 2430 Pulse Mode only)
Source Value, Range, Auto Range
Sense Protection, Range, Auto Range
SYSTem:AZERo:STATe
SYSTem:GUARd
SYSTem:RSENse
ROUTe:TERMinals
CALCulate1:STATe
CALCulate1:MATH[:EXPRession]:NAME
CALCulate2:FEED
CALCulate2:NULL:OFFSet
CALCulate2:NULL:STATe
CALCulate2:LIMit[1]:STATe
CALCulate2:LIMit[1]:COMPliance:FAIL
CALCulate2:LIMit[1]:COMPliance:SOURce2
CALCulate2:LIMitX:STATe
CALCulate2:LIMitX:UPPer[:DATA]
CALCulate2:LIMitX:UPPer:SOURce2
CALCulate2:LIMitX:LOWer[:DATA]
CALCulate2:LIMitX:LOWer:SOURce2
CALCulate2:LIMitX:PASS:SOURce2