User manual

2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-81
50 Set limit to 50V
60 Set limit to 60V
80 Set limit to 80V
81 to 105 Set limit to NONE
DEFault Set limit to 105V (NONE)
MINimum Set limit to 10V
MAXimum Set limit to 105V (NONE)
2440
<n> = -42 to 42 Specify V-Source limit
4 Set limit to 4V
8 Set limit to 8V
12 Set limit to 12V
16 Set limit to 16V
20 Set limit to 20V
24 Set limit to 24V
32 Set limit to 32V
33 to 42 Set limit to NONE
DEFault Set limit to 42V (NONE)
MINimum Set limit to 4V
MAXimum Set limit to 42V (NONE)
Query [:LIMit]? Query limit level
[:LIMit]? DEFault Query *RST default limit
[:LIMit]? MINimum Query lowest allowable limit
[:LIMit]? MAXimum Query highest allowable limit
Description This command is used to set the Over Voltage Protection (OVP) limit
for the V-Source. The V-Source output will not exceed the selected
limit. An exception to this is a parameter value that exceeds 160V for
the Model
2400, 500V for the Model 2410, 48V for the Model 2420,
80V for the Models 2425 and 2430, or 32V for the Model 2440.
Exceeding those values allows the V-Source to output its maximum
voltage. The OVP limit is also enforced when in the I-Source Mode.
The limit parameter values are magnitudes and are in effect for both
positive and negative output voltage. You can express the limit as a
positive or negative value.
If you specify a value that is less than the lowest limit, the lowest limit
will be selected. If you specify a value that is between limits, the
lower limit will be selected. For example, if you specify a value of 110
for the Model 2400, the 100V limit will be selected.