User manual
2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-79
2425 and 2430 DC Mode
<n> = -3.15 to 3.15 Set I-Source amplitude (amps)
-105 to 105 Set V-Source amplitude (volts)
DEFault 0A or 0V
MINimum -3.15A or -105V
MAXimum +3.15A or +105V
2430 Pulse Mode
<n> = -10.5 to 10.5 Set I-Source amplitude (amps)
-105 to 105 Set V-Source amplitude (volts)
DEFault 0A or 0V
MINimum -10.5A or -105V
MAXimum +10.5A or +105V
2440
<n> = -5.25 to 5.25 Set I-Source amplitude (amps)
-42 to 42 Set V-Source amplitude (volts)
DEFault 0A or 0V
MINimum -5.25A or -42V
MAXimum +5.25A or +42V
Query :TRIGgered? Query triggered amplitude for fixed source
:TRIGgered? DEFault Query *RST default amplitude
:TRIGgered? MINimum Query lowest allowable amplitude
:TRIGgered? MAXimum Query highest allowable amplitude
Description This command performs the same as the [:IMMediate][:AMPLitude]
command except that the amplitude is not updated immediately.
With this command, the amplitude is updated when the SourceMeter
is triggered to perform a source-measure operation. For example, if
the instrument is waiting in the trigger layer for an external trigger,
the amplitude of the source will not update until that external trigger
is received by the SourceMeter. See
Section 11, “Trigger models,”
for details on trigger model operation.
The MINimum and MAXimum parameters are only valid if the highest
source range is presently selected. Sending the MINimum or
MAXimum parameters on a lower source range will generate error -
221 (Setting Conflict).
Set voltage limit
[:LEVel] <n>
:SOURce[1]:VOLTage:PROTection[:LEVel] <n> Set voltage limit for V-Source
Parameters 2400
<n> = -210 to 210 Specify V-Source limit
20 Set limit to 20V
40 Set limit to 40V
60 Set limit to 60V