2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-77
Set amplitude for fixed source
[:IMMediate][:AMPLitude] <n>
:SOURce[1]:CURRent[:LEVel][:IMMediate][:AMPLitude] <n> Set fixed I-Source amplitude immedi-
ately
:SOURce[1]:VOLTage[:LEVel][:IMMediate][:AMPLitude] <n> Set fixed V-Source amplitude immedi-
ately
Parameters 2400/2400-LV/2401
<n> = -1.05 to 1.05 Set I-Source amplitude (amps)
-210 to 210 Set V-Source amplitude (volts)
(-21 to 21, 2400-LV and 2401)
DEFault 0A or 0V
MINimum -1.05A or -210V
MAXimum +1.05A or +210V (+21V, 2400-LV
and 2401)
2410
<n> = -1.05 to 1.05 Set I-Source amplitude (amps)
-1100 to 1100 Set V-Source amplitude (volts)
DEFault 0A or 0V
MINimum -1.05A or -1100V
MAXimum +1.05A or +1100V
2420
<n> = -3.15 to 3.15 Set I-Source amplitude (amps)
-63 to 63 Set V-Source amplitude (volts)
DEFault 0A or 0V
MINimum -3.15A or -63V
MAXimum +3.15A or +63V
2425 and 2430 DC Mode
<n> = -3.15 to 3.15 Set I-Source amplitude (amps)
-105 to 105 Set V-Source amplitude (volts)
DEFault 0A or 0V
MINimum -3.15A or -105V
MAXimum +3.15A or +105V
2430 Pulse Mode
<n> = -10.5 to 10.5 Set I-Source amplitude (amps)
-105 to 105 Set V-Source amplitude (volts)
DEFault 0A or 0V
MINimum -10.5A or -105V
MAXimum +10.5A or +105V
2440
<n> = -5.25 to 5.25 Set I-Source amplitude (amps)
-42 to 42 Set V-Source amplitude (volts)
DEFault 0A or 0V
MINimum -5.25A or -42V
MAXimum +5.25A or +42V