User manual

18-70 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
Configure and control filter
NOTE For the Model 2430 Pulse Mode, filtering is not used. Therefore, the follow-
ing filter commands are not valid for the Model 2430 Pulse Mode.
TCONtrol <name>
[:SENSe[1]]:AVERage:TCONtrol <name> Select filter type
Parameters <name> = REPeat Repeating filter
MOVing Moving filter
Query :TCONtrol? Query filter type
Description This command is used to select the type of averaging filter (REPeat
or MOVing). These filter types are explained in
Section 7, Filters.”
The number of readings that are averaged by the filter is set with the
:AVERage:COUNt command. The :AVERage:STATe command is
used to enable or disable the filter.
NOTE For the Model 2430 Pulse Mode, this command is ignored.
COUNt <n>
[:SENSe[1]]:AVERage:COUNt <n> Specify filter count
Parameters <n> = to 100 Specify filter count
DEFault 10
MINimum 1
MAXimum 100
Query :COUNt? Query filter count
:COUNt? DEFault Query the *RST default filter count
:COUNt? MINimum Query the lowest allowable filter count
:COUNt? MAXimum Query the largest allowable filter count
Description These commands are used to specify the filter count. In general, the
filter count is the number of readings that are acquired and stored in
the filter buffer for the averaging calculation. The larger the filter
count, the more filtering that is performed.
NOTE For the Model 2430 Pulse Mode, this command is ignored.