User manual

2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-67
Set compliance parameters
[:LEVel] <n>
[:SENSe[1]]:CURRent[:DC]:PROTection[:LEVel] <n> Set current compliance
[:SENSe[1]]:VOLTage[:DC]:PROTection[:LEVel] <n> Set voltage compliance
Parameters 2400
<n> = -1.05 to 1.05 Current compliance limit
-210 to 210 Voltage compliance limit
DEFault 105uA, 21V
MINimum -1.05A, -210V
MAXimum 1.05A, 210V
2400-LV and 2401
<n> = -1.05 to 1.05 Current compliance limit
-21 to 21 Voltage compliance limit
DEFault 105uA, 21V
MINimum -1.05A, -21V
MAXimum 1.05A, 21V
2410
<n> = -1.05 to 1.05 Current compliance limit
-1100 to 1100 Voltage compliance limit
DEFault 105uA, 21V
MINimum -1.05A, -1100V
MAXimum 1.05A, 1100V
2420
<n> = -3.15 to 3.15 Current compliance limit
-63 to 63 Voltage compliance limit
DEFault 105uA, 21V
MINimum -3.15A, -63V
MAXimum 3.15A, 63V
2425 and 2430 DC Mode
<n> = -3.15 to 3.15 Current compliance limit
-105 to 105 Voltage compliance limit
DEFault 105uA, 21V
MINimum -3.15A, -105V
MAXimum 3.15A, 105V
2430 Pulse Mode
<n> = -10.5 to 10.5 Current compliance limit
-105 to 105 Voltage compliance limit
DEFault 105uA, 21V
MINimum -10.5A, -105V
MAXimum 10.5A, 105V
2440
<n> = -5.25 to 5.25 Current compliance limit
-42 to 42 Voltage compliance limit
DEFault 105uA, 21V
MINimum -5.25A, -42V