User’s Manual SCPI Command Reference 18-57
With NORMal selected, the V-Source is selected and set to 0V when
the output is turned off. Compliance is set to 0.5% full scale of the
present current range.
In the ZERO output-off state when the V-Source OUTPUT is turned
off, the V-Source is set to 0V and current compliance is not changed.
When the I-Source OUTPUT is turned off, the V-Source mode is
selected and set to 0V. Current compliance is set to the programmed
Source I value or to 0.5% full scale of the present current range,
whichever is greater.
The ZERO output-off state is typically used with the V-Source and
Output Auto-On (see the :SOURce1:CLEar:AUTO command) to gen
-
erate voltage waveforms that alternate between 0V and the pro-
grammed output-on voltage.
With GUARd selected, the I-Source is selected and set to 0A. Volt-
age compliance is set to 0.5% full scale of the present voltage range.
This output-off state should be used when performing 6-wire guarded
ohms measurements or for any other load that uses an active
source.
NOTE For more information on output-off states, see Section 13, “Output-off
states.” For the Model 2430 Pulse Mode, the output-off state is always NOR-
Mal.
ROUTe subsystem
The ROUTe subsystem is summarized in Table 18-5.
Select input jacks
TERMinals <name>
:ROUTe:TERMinals <name> Select front or rear panel in/out jacks
Parameters <name> = FRONt Front panel in/out jacks
REAR Rear panel in/out jacks
Query :TERMinals? Query state of front/rear switch setting
Description This command is used to select which set of input/output terminals to
enable (front panel or rear panel).