User manual

2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-51
Bit 17 (Offset Compensation) — Set to 1 if Offset Compensated Ohms is
enabled.
Bit 18 — Contact check failure (see Appendix F).
Bits 19, 20 and 21 (Limit Results) — Provides limit test results
(see grading and sorting modes below).
Bit 22 (Remote Sense) — Set to 1 if 4-wire remote sense selected.
Bit 23 (Pulse Mode) — Set to 1 if in the Pulse Mode.
Description Bits 8, 9, and 19-21 flag pass/fail conditions for the various limits
tests. The bit values for the grading and sorting modes are covered
below. See :CALC2:CLIM:MODE and associated commands in
“Cal-
culate subsystems,” page 18-26.
Sorting mode status bit values:
Result
1
Bit #: 21 20 19 9 8
Meas. Event
Status
2
Limit 1 and 4 pass and 2, 3, and
5-12 disabled
0 0 0 0 0 Bit 5 (LP)
Limit test 1 fail 0 0 0 0 1 Bit 0 (L1)
Limit test 2 pass 0 0 0 1 0 Bit 5 (LP)
Limit test 3 pass 0 0 0 1 1 Bit 4 (HL3)
Limit test 4 fail 0 0 1 0 1 Bit 10 (CC)
Limit test 5 pass 0 0 1 0 0 Bit 5 (LP)
Limit test 6 pass 0 0 1 1 0 Bit 5 (LP)
Limit test 7 pass 0 0 1 1 1 Bit 5 (LP)
Limit test 8 pass 0 1 0 0 0 Bit 5 (LP)
Limit test 9 pass 0 1 0 0 0 Bit 5 (LP)
Limit test 10 pass 0 1 0 1 0 Bit 5 (LP)
Limit test 11 pass 0 1 0 1 1 Bit 5 (LP)
Limit test 12 pass 0 1 1 0 0 Bit 5 (LP)
Limit 1 and 4 pass and 2, 3, and
5-12 fail
1 1 1 1 1 -
1. Limit 4 test available only with contact check option (Appendix F).
2. See Section 15, Measurement Event Register” and Figure 15-6 for details.