18-42 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
the first contact check (optional), compliance, hi, or low failure. Oth-
erwise, the CALC2:CLIM:PASS:SOUR2 pattern will be output.
In SORTing mode, a reading will fail if it fails the contact check test,
the compliance test, or is not within any of the Digital I/O Bands. If
the tests pass and only LIMIT 1 or 4 is enabled,
CALC2:CLIM:PASS:SOUR2 pattern will be output. Otherwise, the
first limit test band that passes will output its LOW:SOUR2 pattern
(UPP:SOUR2 patterns will be ignored). If LIMIT1 or LIMIT 4 fails,
their SOUR2 patterns will be output. If no LIMIT2, 3, 5-12 limit
passes, the CALC2:CLIM:FAIL:SOUR2 pattern will be output.
Clear test results
[:IMMediate]
:CALCulate2:CLIMits:CLEar[:IMMediate] Clears test results and resets Digital I/O Port
Description This command clears the test results (pass or fail) of the limit tests
and resets the output lines of the Digital I/O port back to the
:SOURce2:TTL settings. (See
“SOURce2,” page 18-98, subsystem.)
AUTO <b>
:CALCulate2:CLIMits:CLEar:AUTO <b> Control auto-clear for test results
Parameters <b> = 1 or ON Enable auto-clear
0 or OFF Disable auto-clear
Query :AUTO? Query state of auto-clear
Description With auto-clear enabled, test results will clear and the output lines of
the Digital I/O port will reset when the :INITiate command is sent to
start a new test sequence.
When disabled, you must use :IMMediate to perform the clear
actions.