User’s Manual SCPI Command Reference 18-35
NOTE See Appendix C, “Data Flow,” for a detailed explanation on how data flows
through the various operation blocks of the SourceMeter. It clarifies the type
of readings that are acquired by the various commands to read data.
LATest?
:CALCulate2:DATA:LATest? Read latest CALC2 data
Description This command works exactly like CALC2:DATA, except it returns
only the latest null offset or limit result.
Configure and control limit tests
COMPliance:FAIL <name>
:CALCulate2:LIMit[1]:COMPliance:FAIL <name>
Parameters <name> = IN Fail Limit 1 test when unit goes
into compliance
OUT Fail Limit 1 test when unit comes
out of compliance
Query :FAIL? Query when Limit 1 test failure occurs
Description This command is used to specify the condition that will cause Limit 1
test to fail. With IN specified, the test will fail when the SourceMeter
goes into compliance. With OUT specified, the test will fail when the
SourceMeter comes out of compliance.
[:DATA] <n>
:CALCulate2:LIMitx:LOWer[:DATA] <n> Specify lower LIMIT x (x = 2, 3, 5-12)
:CALCulate2:LIMitx:UPPer[:DATA] <n> Specify upper LIMIT x (x = 2, 3, 5-12)
Parameters <n> = -9.999999e20 to Specify limit value
9.999999e20
DEFault Set specified lower limit to -1
Set specified upper limit to 1
MINimum Set specified limit to -9.999999e20
MAXimum Set specified limit to +9.999999e20
Query :UPPer? Query specified upper limit
:UPPer? DEFault Query *RST default upper limit
:UPPer? MINimum Query lowest allowable upper limit
:UPPer? MAXimum Query largest allowable upper limit
:LOWer? Query specified lower limit
:LOWer? DEFault Query *RST default lower limit
:LOWer? MINimum Query lowest allowable lower limit
:LOWer? MAXimum Query largest allowable lower limit