18-18 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
[:SOURce[1]]
:SWEep Configure SWEep source mode: ✓
:SPACing <name> Select sweep spacing type (LINear or
LOGarithmic).
LINear ✓
:SPACing? Query sweep spacing. ✓
:POINts <n> Specify number of sweep points (2 to
2500).
2500 ✓
:POINts? Query number of points in sweep. ✓
:DIRection <name> Sweep from start to stop (UP) or from stop
to start (DOWN).
UP
:DIRection? Query sweep direction.
:RANGing <name> Select source ranging mode (BEST,
AUTO, or FIXed).
BEST ✓
:RANGing? Query source ranging mode. ✓
:CABort <name> Abort on compliance (NEVer, EARLy, or
LATE).
NEVer
:CABort? Query abort on compliance state.
:LIST Configure LIST source mode: ✓
:CURRent <NRf> Create list of I-Source values.
1
:APPend <NRf> Add I-Source values to end of list. ✓
:POINts? Query number of source values in list.
:STARt <n> Specify list sweep start point. 1
:STARt? Query list sweep start point.
:CURRent? Query I-Source list. ✓
:VOLTage <NRf> Create list of V-Source values. 1
:APPend <NRf> Add V-Source values to end of list.
:POINts? Query number of source values in list.
:STARt <n> Specify list sweep start point. 1
:STARt? Query list sweep start point.
:VOLTage? Query V-Source list.
:MEMory Configure Source Memory Sweep:
:SAVE <n> Save settings at memory location (1 to 100).
:RECall <n> Recall settings from memory (1 to 100). 1
:POINts <n> Specify number of sweep points (1 to 100).
:POINts? Query number of sweep points.
:STARt <NRf> Specify start location for Source Memory
Sweep (1 to 100).
1
:STARt? Query start location.
1. The parameter values for this command are provided with the detailed command reference information. Detailed infor-
mation on SCPI commands follows the last command summary table in this section.
Table 18-7 (continued)
SOURce command summary
Command Description
Default
parameter SCPI
Source
memory