User manual

2400 Series SourceMeter
®
User’s Manual Status Structure 15-15
Questionable Event Register
The used bits of the Questionable Event Register (shown in Figure 15-7) are
described as follows:
Bits B0 through B7 — Not used.
Bit B8, Calibration Summary (Cal) — Set bit indicates that an invalid cali-
bration constant was detected during the power-up sequence. This error
will clear after successful calibration of the instrument.
Bits B9 through B13 — Not used.
Bit B14, Command Warning (Warn) — Set bit indicates that a Signal Ori-
ented Measurement Command parameter has been ignored.
Bit B15 — Not used.
Figure 15-7
Questionable event status
Questionable
Event Enable
Register
&
&
OR
To QSB of Status
Byte Register.
(See Figure 15-3.)
Warn = Command Warning
Cal = Calibration Summary
& = Logical AND
OR = Logical OR
(B7 - B0)
CAL
(B8)
(B13 - B9)
(B15)
Warn
(B14)
Questionable
Condition
Register
Questionable
Event Register
(B7 - B0)
CAL
(B8)
(B13 - B9)
(B15)
Warn
(B14)
(B7 - B0)
CAL
(B8)
(B13 - B9)
(B15)
Warn
(B14)
:stat:ques:cond?
:stat:ques?
:stat:ques:enab<NRf>
stat:ques:enab?