User manual
15-14 Status Structure 2400 Series SourceMeter
®
User’s Manual
• Bit B7, Reading Overflow (ROF) — Set bit indicates that the volts or amps
reading exceeds the selected measurement range of the SourceMeter.
• Bit B8, Buffer Available (BAV) — Set bit indicates that there are at least two
readings in the buffer.
• Bit B9, Buffer Full (BFL) — Set bit indicates that the trace buffer is full.
• Bit B10 — Limit 4 - Contact Check (see Appendix F).
• Bit B11, Output Enable Asserted (Int) — Set bit indicates that the output
enable line is at digital low (asserted). The source output can be turned on.
• Bit B12, Over Temperature (OT) — Set bit indicates that an over tempera-
ture condition exists. The source output cannot be turned on.
• Bit B13, Over Voltage Protection (OVP) — Set bit indicates that the source
is being limited at the programmed limit level.
• Bit B14, Compliance (Comp) — Set bit indicates that the source is in com-
pliance.
• Bit B15 — Not used.
Figure 15-6
Measurement event status
LL3
(B3)
HL2
(B2)
LL2
(B1)
L1
(B0)
HL3
(B4)
LP
(B5)
RAV
(B6)
ROF
(B7)
BAV
(B8)
BFL
(B9)
CC
(B10)
OT
(B12)
Measurement
Condition Register
BFL
(B9)
CC
(B10)
Measurement Event
Register
CC
(B10)
Measurement Event
Enable Register
&
&
&
&
&
&
OR
To Measurement
Summary Bit
(MSB) of Status
Byte Register.
(See Figure 15-3.)
Comp = In Compliance
OVP = Over Voltage protection acgentcy
OT = Over temperature
OE = Output Enble Asserted
CC = Contact Check
BFL = Buffer Full
BAV = Buffer Available
ROF = Reading Overflow
RAV = Reading Available
OE
(B11)
&
&
&
&
LP = Limits Pass
HL3 = High Limit 3
LL3 = Low Limit 3
HL2 = High Limit 2
LL2 = Low Limit 2
L1 = Limit 1
& = Logical AND
OR = Logical OR
OVP
(B13)
Comp
(B14)
—
(B15)
OE
(B11)
OT
(B12)
OVP
(B13)
Comp
(B14)
—
(B15)
OE
(B11)
OT
(B12)
OVP
(B13)
Comp
(B14)
—
(B15)
&
&
&
&
stat:meas:cond?
stat:meas?
stat:meas:enab <NRF>
stat:meas:enab?
&
L1
(B0)
LL2
(B1)
HL2
(B2)
LL3
(B3)
HL3
(B4)
LP
(B5)
RAV
(B6)
ROF
(B7)
BAV
(B8)
BFL
(B9)
BAV
(B8)
ROF
(B7)
RAV
(B6)
LP
(B5)
HL3
(B4)
LL3
(B3)
HL2
(B2)
LL2
(B1)
L1
(B0)