User manual

11-24 Triggering 2400 Series SourceMeter
®
User’s Manual
Output triggers
All three output triggers are available in the trigger layer for the Pulse Mode. If
enabled, the output triggers occur where indicated in the trigger model.
Invalid trigger settings
Front panel operation
For front panel operation, most trigger settings are done from the “CONFIGURE
TRIGGER” menu which is accessed by pressing CONFIG and then TRIG. For
Pulse Mode operation, the following settings for the “TRIG LAYER” are ignored.
DELAY — Trigger delay is not used. Therefore, the trigger DELAY setting is
ignored.
TRIGGER OUT EVENTS — For the DC Mode, source delay is set from the
source configuration menu, which is accessed by pressing CONFIG and then
SOURCE V or I. When the Pulse Mode is selected, the source delay menu items
(DELAY and AUTO DELAY) are replaced by the pulse DELAY and PULSE
WIDTH menu items.
Remote operation
The following commands are not valid in the Pulse Mode and are therefore
ignored:
:TRIGger:DELay — Trigger delay is not used in the Pulse Mode.
:SOURce:DELay and :SOURce:DELay:AUTO — Source delay is not used in the
Pulse Mode.