User manual
2400 Series SourceMeter
®
User’s Manual Sweep Operation 10-23
Source memory sweep programming example
Table 10-8 summarizes the basic remote command sequence for performing the
basic source memory sweep.
Table 10-8
Source memory sweep programming example
Command Description
*RST
:SENS:FUNC:CONC OFF
:SOUR:FUNC MEM
:SOUR:MEM:POIN 3
:SOUR:MEM:STAR 1
:SOUR:FUNC VOLT
:SENS:FUNC ‘CURR:DC’
:SOUR:VOLT 10
:SOUR:MEM:SAVE 1
:SOUR:FUNC CURR
:SENS:FUNC ‘VOLT:DC’
:SOUR:CURR 100E-3
:SOUR:MEM:SAVE 2
:SENS:FUNC ‘CURR:DC’
:SOUR:MEM:SAVE 3
:TRIG:COUN 3
:OUTP ON
:READ?
Restore GPIB default conditions.
Turn off concurrent functions.
Source memory sweep mode.
Number memory points = 3.
Start at memory location 1.
Volts source function.
Current sense function.
10V source voltage.
Save in source memory location 1.
Current source function.
Volts sense function.
100mA source current.
Save in source memory location 2.
Current sense function.
Save in source memory location 3.
Trigger count = # sweep points.
Turn on source output.
Trigger sweep, request data.