Sweep branching
When using a Source Memory Sweep while performing limit tests, the normal
sequence of sweep memory points can be changed. This is useful when, based
on the results of an initial test, a different set of tests are needed.
The sweep can branch to a specified memory location point, or proceed to the
next memory location in the list. When a memory location is specified, the sweep
will branch to that memory location if the test is successful (PASS condition). If not
successful (FAIL condition), the sweep proceeds to the next memory location in
the list. With NEXT selected (the default), the sweep proceeds to the next memory
location in the list regardless of the outcome of the test (PASS or FAIL condition).
Figure 10-3 shows a six-point sweep branching example. In this case, the unit is
programmed to branch to location 7 when a pass conditions occurs at location 3.
Mode Remote command
Limit 1 bit pattern
Enable/disable Limit X**
Limit X upper limit
Limit X upper bit pattern
Limit X lower limit
Limit X lower bit pattern
Composite limits bit pattern
Next pass memory location
Trigger delay†
Guard type
CALCulate2:LIMit[1]:COMPliance:SOURce2
CALCulate2:LIMitX:STATe
CALCulate2:LIMitX:UPPer[:DATA]
CALCulate2:LIMitX:UPPer:SOURce2
CALCulate2:LIMitX:LOWer[:DATA]
CALCulate2:LIMitX:LOWer:SOURce2
CALCulate2:CLIMits:PASS:SOURce2
CALCulate2:CLIMits:PASS:SMLocation
TRIGger:DELay
SYSTem:GUARd
* X = CURRent or VOLTage.
** Limit X = Limit 2, 3, 5-12.
Not available when the Model 2430 is in the Pulse Mode.
Table 10-2 (cont.)
Source memory saved configurations