User manual

10-8 Sweep Operation 2400 Series SourceMeter
®
User’s Manual
Restoring source memory setups
In addition to automatically sweeping through source memory locations (see
Configuring and running a sweep” later in this section), you can also recall them
individually as follows:
1. Press MENU to display the MAIN MENU.
2. Select SAVESETUP, then press ENTER.
3. Choose SOURCE MEMORY, then press ENTER.
4. Select RESTORE, then press ENTER.
5. Select the source memory location to restore (1-100), then press ENTER.
Saving multiple source memory sweeps
If desired, you can save multiple source memory sweeps in the 100 memory loca-
tions. For example, you could save setups in locations 1 through 4 for one sweep,
and other setups in any other range of memory locations such as locations 50
through 58. To select which sweep to execute, simply select two settings: (1) the
sweep start location, and (2) the number of sweep points. (See “Performing a
source memory sweep” later in this section.)
Saved source memory configurations
Table 10-2 summarizes the configurations that are saved at each source memory
location along with the equivalent remote command. See Section 18 for more
details on these remote commands. The SCPI command reference tables, Table
18-1 through Table 18-11, also list source memory parameters.