6-36 Source-Measure Concepts 2400 Series SourceMeter
®
User’s Manual
Buffer considerations
When the SourceMeter is in the process of storing readings, configuration
changes affect what gets stored in the buffer. These storage considerations and
restrictions are summarized in Table 6-3.
The first column of Table 6-3 shows the SourceMeter configuration when the stor-
age process is started. The next three columns show what happens when config-
uration changes are made while the SourceMeter is storing readings.
Changing V, I, or measurement function
If you started with only a basic measurement function selected, the buffer
will track a basic measurement function change. For example, if you started
in volts and changed to current, the buffer will store the current readings.
If you started with MATH, REL, and/or LIMITS enabled, the buffer will stop
storing readings if you change the basic measurement function. Storage
will continue if you return to the original configuration.
See Section 3, “Basic source-measure procedure,” for more information on
selecting the measurement function.
Table 6-3
Buffer considerations
Configuration at start of
the storage process
What happens if
the basic measure-
ment function (V, I,
or ) is changed?
What happens if
the MATH function
is changed?
What happens if REL or
LIMITS is changed?
Measure V, I, or
MATH (FCTN) enabled
REL/LIMITS enabled
Buffer tracks
Buffer pauses
Buffer pauses
V, I, or stored
MATH not stored
OK
Buffer pauses
V, I, or stored
REL or Limits not stored
MATH stored
REL or Limits not stored
OK