User’s Manual Source-Measure Concepts 6-27
Guard
WARNING GUARD is at the same potential as output HI. Thus, if hazard-
ous voltages are present at output HI, they are also present at
the GUARD
terminal.
NOTE See Section 2, “Guarding methods,” for details on guarded test connec-
tions as well as information on selecting guard modes.
For 6-wire ohms guard measurements, use the GUARD output-off state.
For details on the GUARD output-off state, see
Section 13, “Output-off
states.”
The driven guard (available at the rear panel GUARD terminal) is always enabled
and provides a buffered voltage that is at the same level as the Input/Output HI (or
Sense HI for remote sense) voltage. The purpose of guarding is to eliminate the
effects of leakage current (and capacitance) that can exist between input/output
high and low. In the absence of a driven guard, leakage in the external test circuit
could be high enough to adversely affect the performance of the SourceMeter.
Leakage current can occur through parasitic or non-parasitic leakage paths. An
example of parasitic resistance is the leakage path across the insulator in a coax
or triax cable. An example of non-parasitic resistance is the leakage path through
a resistor that is connected in parallel to the DUT.
There are two programmable output impedance levels for the guard output. The
high-impedance (~10k) CABLE guard is used to reduce the effects of capacitance
and leakage current paths in the test circuit. The low-impedance (<1) OHMS
guard is used to cancel the effects of parallel resistances when measuring a resistor
element of a resistor network.
Cable guard
The CABLE guard selection provides a high-impedance (~10k) driven guard to
prevent positive feedback, which could cause oscillations when using shielded
cables. Cable guard is used to drive the shields of cables and test fixtures. Guard
is extended to a test fixture using a safety banana plug (such as the Model 8008-
BAN). Inside the test fixture, the guard can be connected to a guard plate or shield
that surrounds the DUT.