User manual

6-24 Source-Measure Concepts 2400 Series SourceMeter
®
User’s Manual
Figure 6-14
Source I
Source V
When configured to source voltage (V-Source) as shown in Figure 6-15, the
SourceMeter functions as a low-impedance voltage source with current limit capa-
bility and can measure current (I-Meter) or voltage (V-Meter).
Sense circuitry is used to continuously monitor the output voltage and make
adjustments to the V-Source as needed. The V-Meter senses the voltage at the
Input/Output terminals (2-wire local sense) or at the DUT (4-wire remote sense
using the sense terminals) and compares it to the programmed voltage level. If the
sensed level and the programmed value are not the same, the V-Source is
adjusted accordingly. Remote sense eliminates the effect of voltage drops in the
test leads ensuring that the exact programmed voltage appears at the DUT.
NOTE The voltage error feedback to the V-Source is an analog function. The
source error amplifier is used to compensate for IR drop in the test leads.
Remote
SENSE HI
IN/OUT HI
Remote
IN/OUT LO
I-Source
x1
V-Meter
SENSE LO
Local
Local
GUARD SENSE
GUARD
+
I-Meter