User manual

2400 Series SourceMeter
®
User’s Manual Source-Measure Concepts 6-5
measure range = real compliance). On the 20V and 200mV measurement ranges,
compliance is 21V and 210mV, respectively (measure range < compliance setting
= range compliance). The same rules apply for the next three entries for current
compliance.
Table 6-2
Compliance examples
Compliance setting Measurement range Actual compliance
Display message Setting Display message Range Value Type
Cmpl: 0.15000kV
Cmpl: 0.15000kV
Cmpl: 0.15000kV
Cmpl: 075.000 mA
Cmpl: 075.000 mA
Cmpl: 075.000 mA
150V
150V
150V
75mA
75mA
75mA
---.---V
--.----V
---.---mV
---.---mA
--.----mA
-.-----mA
200V
20V
200mV
100mA
10mA
1mA
150V
21V
210mV
75mA
10.5mA
1.05mA
Real
Range
Range
Real
Range
Range