User’s Manual Pulse Mode Operation (Model 2430 only) 5-11
4. For continuous pulse output, select INFINITE, press ENTER and proceed
to step
6. Otherwise, select FINITE, press ENTER and proceed to the next
step to set the arm count.
5. Enter the desired arm count value and press ENTER.
6. Use the EXIT key to back out of the menu.
Perform the following steps to set the trigger count:
1. Press CONFIG and then TRIG to display the trigger configuration menu.
2. Select TRIG LAYER from the displayed choices and press ENTER.
3. Select COUNT from the displayed choices and press ENTER.
4. Enter the desired trigger count value and press ENTER.
5. Use the EXIT key to back out of the menu.
Disable/enable auto zero
Pulse speed can be increased by disabling auto zero. However, zero drift will
eventually corrupt the accuracy of the pulse measurement. To re-establish an
accurate pulse measurement, enable auto zero or change the speed setting.
1. Press MENU to display the main menu.
2. Select A/D CTRL from the displayed choices and press ENTER.
3. Select AUTO ZERO from the displayed choices and press ENTER.
4. Select DISABLE or ENABLE and press ENTER.
5. Use the EXIT key to back out of the menu.
Remote command Pulse Mode configuration
For remote operation, the commands to select and configure the Pulse Mode are
provided in Table 18-6. and Table 18-7.