User Manual
2400 Series SourceMeter
®
User’s Manual Ohms Measurements 4-11
Ohms source readback
NOTE For the Model 2430 Pulse Mode, ohms source readback cannot be
enabled.
With ohms source readback enabled, the instrument measures the actual source
value instead of the programmed value used for ohms measurements and then
uses that measured value for reading calculations. Normally, ohms source read-
back should be left enabled for optimum ohms measurement accuracy because
SourceMeter measurement accuracy is better than source programming accu-
racy. (See specifications in Appendix A.) However, disabling source readback will
allow you to make valid ohms measurements with the source in compliance. See
Section 6, “Source I measure I and source V measure V,” for more details.
Ohms source readback selection
Use the following procedure to enable or disable ohms source readback:
1. Press CONFIG then Ω.
2. Select SRC RDBK, then press ENTER.
3. Select DISABLE or ENABLE as desired, then press ENTER.
4. Press EXIT to return to normal display.
NOTE Readings in the compliance field will be invalid with source readback dis-
abled.