User Manual

2400 Series SourceMeter
®
User’s Manual Ohms Measurements 4-5
WARNING To prevent electric shock, do not make or break connections to
the SourceMeter with the output on. If on, press the ON/OFF
OUTPUT key to turn the output off.
1. Select ohms measurement function by pressing MEAS .
2. Select the ohms measurement method (AUTO or MANUAL).
NOTE Use the manual ohms mode and the V-source method when high-speed
settling is required.
3. For manual ohms, configure source.
For manual ohms, you can Source I or Source V at the user-programmed
output level. The lowest allowable compliance limit is based on the load
and the source value. For example, if sourcing 1V to a 1k resistor, the
lowest allowable current compliance is 1mA (1V/1k = 1mA). Setting a limit
lower than 1mA will place the source in compliance.
4. Select measurement range.
Use the RANGE keys to select a range appropriate for the expected ohms
reading, or use autorange by pressing AUTO. When using manual ranging,
selecting the most sensitive (lowest) range provides the best accuracy.
Autorange automatically goes to the most sensitive range.
5. Turn output on.
Turn the output on by pressing the ON/OFF OUTPUT key. The OUTPUT
indicator will turn on to indicate the output is on.
6. Observe reading on display.
The SourceMeter will go into compliance if you exceed the maximum ohms
measurement range.
7. Turn output off.
When finished, turn the output off by pressing the ON/OFF OUTPUT key.
The OUTPUT indicator light will turn off.