2400 Series SourceMeter
®
User’s Manual Basic Source-Measure Operation 3-21
Remote command measure only
Table 3-7 summarizes the basic command sequence for measure only. The steps
outlined correspond to those in the Front panel measure only” sequence above.
Table 3-7
Measure only programming example
Step
1
Action Commands
2,3
Comments
1
2
3
5
6
7
Select measure, source functions
Set source and compliance
Select volts measure range
Turn on output
Read data
Turn off output
*RST
:SOUR:FUNC CURR
:SOUR:CURR:MODE FIXED
:SENS:FUNC “VOLT”
:SOUR:CURR:RANG MIN
:SOUR:CURR:LEV 0
:SENS:VOLT:PROT 25
:SENS:VOLT:RANG 20
:FORM:ELEM VOLT
:OUTP ON
:READ?
:OUTP OFF
Restore GPIB defaults.
Current source function.
Fixed current source mode.
Volts measure function.
Lowest source range.
0A source level.
25V compliance.
20V range.
Volts only.
Output on before measuring.
Trigger, acquire reading.
Output off after measuring.
1. Steps correspond to front panel steps listed previously in “Front panel measure only.”
2. Commands must be sent in order given.
3. Instrument must be addressed to talk after :READ? to acquire data.