2400 Series SourceMeter
®
User’s Manual Basic Source-Measure Operation 3-13
Source delay
NOTE For the Model 2430 Pulse Mode, source delay is not used. The following
information assumes that the Model 2430 is in the DC Mode of operation
(“Vpls” or “Ipls” displayed). To select the DC Mode, press CONFIG and
then SOURCE V or I, select SHAPE from the available menu items, and
then select DC.
The source delay options are used to set the settling time for the source. This
source delay is the delay phase of the Source-Delay-Measure (SDM) cycle. See
Section 6. The auto delay option is used to automatically set the delay. The delay
period is range dependent (Table 3-4).
The delay option is used to manually set the delay from 000.00000 seconds to
9999.99900 seconds. Manually setting the delay disables auto delay.
Front panel source delay
To set the manual source delay from the front panel:
1. Press CONFIG then SOURCE V.
2. Select DELAY from the displayed choices, then press ENTER.
3. Enter the desired DELAY value, then press ENTER.
4. Press EXIT to return to normal display.
To set the auto source delay state from the front panel:
1. Press CONFIG then SOURCE V.
2. Select AUTO DELAY from the displayed choices, then press ENTER.
3. Select ENABLE or DISABLE as desired, then press ENTER.
Table 3-4
Auto source delay
2400/2400-LV/2401
I-range
2410
I-range
2420
I-range
2425/2430
I-range
2440
I-range
Auto delay
(Source V)
Auto
delay
(Source I)
1µA
10µA
100µA
1mA
10mA
100mA
1A
1µA
10µA
100µA
1mA
20mA
100mA
1A
10µA
100µA
1mA
10mA
100mA
1A
3A
10µA
100µA
1mA
10mA
100mA
1A
3A/10A*
10µA
100µA
1mA
10mA
100mA
1A
5A
3msec
2msec
1msec
1msec
1msec
1msec
1msec
3msec
1msec
1msec
1msec
1msec
1msec
2msec
*10A range only on Model 2430.