2-12 Connections 2400 Series SourceMeter
®
User’s Manual
Sense and guard selections
NOTE When sense or guard settings are changed, the OUTPUT will turn OFF.
Sense selection
When using the SENSE HI and LO terminals of the SourceMeter, 4-wire remote
sensing must be selected. When not using these terminals, local sensing must be
selected. Sense selection is explained as follows. See Section 3, “Operation over-
view,” for details on sense.
NOTE The SourceMeter defaults to the 2-wire (local) sense mode when the out-
put is off, regardless of the sense setting. When the output is turned on,
the present sense setting will be in effect.
Front panel sense selection
On power-up, the instrument is automatically set for 2-wire local sense. Perform
the following steps to check and/or change the sense selection:
1. Press CONFIG and then the SOURCE V, MEAS V, or key. Changing
sense in one configuration menu changes it in all of the others.
2. For SOURCE V and , use the ß and © keys to place the cursor (flashing
menu item) on SENSE MODE and press ENTER.
NOTE Cursor position indicates the present sense selection. Two-wire indicates
that local sense is selected, and 4-wire indicates that remote sense is
selected. To retain the present selection, use the EXIT key to back out of
the menu structure and skip the next two steps.
3. To change the sense selection, place the cursor on 4-wire to select remote
sense or on 2-wire to select local sense, and press ENTER.
4. Use the EXIT key to back out of the menu structure.
Remote command sense selection
Use the :SYSTem:RSENse command (Section 18, “SYSTem subsystem”) to
choose between local and remote sensing via remote. For example, send the fol-
lowing command to enable remote sensing:
:SYST:RSEN ON
Conversely, send this command to disable remote sensing (enable local sensing):
:SYST:RSEN OFF