2400 Series SourceMeter
®
User’s Manual Connections 2-9
Figure 2-4
High-impedance guarding (cable guard)
Ohms guard
Use the guarded ohms connection schemes shown in Figure 2-5 for the following
source-measure operation:
In-circuit resistance measurements on the DUT where other parasitic leak-
age devices are present.
Note that ohms guard must be selected for this connection scheme. Ohms guard
is not available for the 1A range (source or measure). See “Guard selection,”
page 2-13, to select ohms guard and Section 4, “6-wire ohms measurements.”
INPUT/
OUTPUT
4-WIRE
SENSE
HI
LO
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
V,
GUARD
GUARD
SENSE
WARNING: NO INTERNAL OPERATOR SERVICA
CAUTION: FOR CONTINUED PROTECTION AGAINST FIR
RS232
DUT
Connect to earth safety ground
using #18 AWG wire or larger.
Guard Shield Test Fixture