User Manual
2400 Series SourceMeter
®
User’s Manual Contact Check Function F-15
Contact check programming example
Tab l e F - 4 summarizes a basic contact check programming example.
Command Description
:SYSTem:RSENse ON Enable 4-wire remote sense for contact check.
:SYSTem:RSENse? Query sense selection.
:SYSTem:CCHeck ON | OFF Enable/disable contact checking.
:SYSTem:CCHeck? Query state of contact check test.
:SYSTem:CCHEck:RESistance <NRf> Select threshold resistance (0 to 60):
<NRf> = 0 to 2 Selects 2W threshold resistance
= >2 to 15 Selects 15W threshold resistance
= >15 to 60 Selects 50W threshold resistance
:SYSTem:CCHeck:RESistance? Query threshold resistance.
:TRIGger:SEQuence2:SOURce <name> Enable or disable contact check event detection; <name> =
CCHeck (enable) or IMMediate (disable).
:TRIGger:SEQuence2:SOURce? Query state of event detection.
:TRIGger:SEQuence2:TOUT <NRf> Specify contact check timeout in seconds (0.00000 to
999.99990).
:TRIGger:SEQuence2:TOUT? Query contact check timeout setting.
*RST and SYSTem:PRESet parameter defaults are shown in boldface.
Table F-3 (cont.)
Contact check remote commands