User Manual

2400 Series SourceMeter
®
User’s Manual Contact Check Function F-9
Figure F-5
Sorting mode contact check limit testing
Perform
Limit 1 Test
?
Yes
No
Pass
?
No
Yes
Display
FAIL
Perform
Limit 2,3,5-12
Tests
?
Yes
No
Pass
?
No
Yes
Start
Turn Output ON and press LIMIT key.
Output
Fail Pattern
Test
Another
Device
?
No
Yes
Stop
Press LIMIT
Wait for SOT pulse
from handler
Perform Source-
Measure action
Display
PASS
Output Limit 1
Fail Pattern
Output Limit 2,3,5-12
Pass Pattern
Display
FAIL
Limits
2, 3 and 5-12
Disabled
?
Display
PASS
Output Pass
Pattern
Yes
No
Limits
1-3 and 5-12
Disabled
?
Yes
No
Display
PASS
Output Pass
Pattern
Pass
?
Yes
Display
FAIL
Output Limit 4
Fail Pattern
No
Perform
Limit 4 Test
?
Yes
No
Perform
Contact Check
Enabling contact check does not automatically enable LIMIT 4. LIMIT 4
must be explicitly enabled from the front panel or using the :CALC2 subsystem com-
mands. SeeFront panel contact check operation, page F-11, or Remote contact
check operation, page F-14 for more details.