User Manual
18-122 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
NOTE Each event in the list must be separated by a comma. (i.e., :arm:output
source, delay, sense).
Query :OUTPut? Query output trigger event(s)
Description This command is used to specify when trigger pulses occur on the
specified output trigger line of the Trigger Link (see
“OLINe <NRf>,”
page 18-121).
Arm Layer Triggers — With TEXit selected, an output trigger will
occur when exiting the trigger layer. With TENTer selected, an output
trigger will occur when entering the trigger layer. With NONE
selected, the arm layer output trigger is disabled.
Trigger Layer Triggers — You can specify from one to all three
events. Each event in the list must be separated by a comma (,).
The SOURce, DELay, and MEASure events refer to the Source-
Delay-Measure (SDM) cycle. This is the Device Action in the Trigger
Model. (See
Figure 11-8 and Figure 11-9.) With SOURce specified,
an output trigger occurs after the source is set. With DELay speci-
fied, an output trigger occurs after the delay period. With MEASure
specified, an output trigger occurs after the measurement.
NOTE When disabling triggers, the NONE parameter must be sent alone (i.e.,
trig:outp none). If it is listed with any of the other event parameters, NONE
will be ignored.