User Manual
2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-121
The Delay Event Detector will be disabled since the DELay parame-
ter is not included in the above event list.
NOTE To disable all the Trigger Layer event detectors, the NONE parameter must
be sent alone (i.e. trigger:input none). If it is listed with any of the other
parameters, NONE will be ignored.
ILINe <NRf>
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:ILINe <NRf> Select input line; arm layer
:TRIGger[:SEQuence[1]][:TCONfigure]:ILINe <NRf> Select input line; trigger layer
Parameters <NRf> = 1 Line #1
2 Line #2
3 Line #3
4 Line #4
Query :ILINe? Query input trigger line
Description This command is used to select input lines for the Trigger Link. For
normal operation, Trigger Link input and output (see “
OLINe <NRf>”)
should not share the same line.
OLINe <NRf>
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:OLINe <NRf> Select output line; arm layer
:TRIGger[:SEQuence[1]][:TCONfigure]:OLINe <NRf> Select output line; trigger layer
Parameters <NRf> = 1 Line #1
2 Line #2
3 Line #3
4 Line #4
Query :OLINe? Query output trigger line
Description This command is used to select output lines for the Trigger Link. For
normal operation, Trigger Link input and output (see “
ILINe <NRf>”)
should not share the same line.
OUTPut <event list>
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:OUTPut <event list> Arm layer events
:TRIGger[:SEQuence[1]][:TCONfigure]:OUTPut <event list> Trigger layer events
Parameters Arm Layer Triggers
<event list >: TENTer Trigger on entering trigger layer
TEXit Trigger on exiting trigger layer
NONE Disable arm layer output trigger
Trigger Layer Triggers
<event list>: SOURce Output trigger after source level is set
DELay Output trigger after delay period
SENSe Output Trigger after measurement
NONE Disable trigger layer triggers