2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-91
Configure list
CURRent <NRf list>
VOLTage <NRf list>
:SOURce[1]:LIST:CURRent <NRf list> Define I-Source list
:SOURce[1]:LIST:VOLTage <NRf list> Define V-Source list
Parameters <NRf list> = NRf, NRf … NRf
2400/2400-LV/2401
NRf = -1.05 to 1.05 I-Source value
-210 to 210 V-Source value (-21 to 21,
2400-LV and 2401)
2410
NRf = -1.05 to 1.05 I-Source value
-1100 to 1100 V-Source value
2420
NRf = -3.15 to 3.15 I-Source value
-63 to 63 V-Source value
2425 and 2430 DC Mode
NRf = -3.15 to 3.15 I-Source value
-105 to 105 V-Source value
2430 Pulse Mode
NRf = -10.5 to 10.5 I-Source value
-105 to 105 V-Source value
2440
NRf = -5.25 to 5.25 I-Source value
-42 to 42 V-Source value
Query :CURRent? Query I-Source list
:VOLTage? Query V-Source list
Description These commands are used to define a list of source values (up to
100) for the list sourcing mode of operation. When operation is
started, the instrument will sequentially source each current or volt
-
age value in the list. A measurement is performed at each source
level.
The following command shows the proper format for defining an I-
Source list using current source values of 10mA, 130mA and 5mA:
:SOURce[1]:LIST:CURRent 0.01, 0.13, 0.005
If using manual source ranging, you can use auto range for source
values that are not within a single range.
NOTE In order to execute a source list, the selected source must be in the list
sourcing mode, and the product of the arm and trigger count should be at
least the same as the number of source points in the list.