18-86 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
CENTer <n>
SPAN <n>
:SOURce[1]:CURRent:CENTer <n> Specify center point of current sweep
:SOURce[1]:VOLTage:CENTer <n> Specify center point of voltage sweep
:SOURce[1]:CURRent:SPAN <n> Specify span of the current sweep
:SOURce[1]:VOLTage:SPAN <n> Specify span of the voltage sweep
Parameters 2400/2400-LV/2401
<n> = -2.1 to 2.1 Set I-Source level (amps)
-420 to 420 Set V-Source level (volts)
(-42 to 42, 2400-LV and 2401)
DEFault 0A or 0V
MINimum -2.1A or -420V (-42V, 2400-LV
and 2401)2400-LV and 2401)
MAXimum +2.1A or +420V (+42V, 2400-LV)
2410
<n> = -2.1 to 2.1 Set I-Source level (amps)
-2200 to 2200 Set V-Source level (volts)
DEFault 0A or 0V
MINimum -2.1A or -2200V
MAXimum +2.1A or +2200V
2420
<n> = -6.3 to 6.3 Set I-Source level (amps)
-128 to 128 Set V-Source level (volts)
DEFault 0A or 0V
MINimum -6.3A or -128V
MAXimum +6.3A or +128V
2425 and 2430 DC Mode
<n> = -6.3 to 6.3 Set I-Source level (amps)
-210 to 210 Set V-Source level (volts)
DEFault 0A or 0V
MINimum -6.3A or -210V
MAXimum +6.3A or +210V
2430 Pulse Mode
<n> = -21 to 21 Set I-Source level (amps)
-210 to 210 Set V-Source level (volts)
DEFault 0A or 0V
MINimum -21A or -210V
MAXimum +21A or +210V
2440
<n> = -10.5 to 10.5 Set I-Source level (amps)
-84 to 84 Set V-Source level (volts)
DEFault 0A or 0V
MINimum -10.5A or -42V
MAXimum +10.5A or +42V