User Manual
18-84 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
on a linear scale. With LOGarithmic selected, the source-measure
points will be performed on a logarithmic scale.
STARt <n>
STOP <n>
:SOURce[1]:CURRent:STARt <n> Specify start current level (current sweep)
:SOURce[1]:VOLTage:STARt <n> Specify start voltage level (voltage sweep)
:SOURce[1]:CURRent:STOP <n> Specify stop current level (current sweep)
:SOURce[1]:VOLTage:STOP <n> Specify stop voltage level (voltage sweep)
Parameters 2400/2400-LV/2401
<n> = -1.05 to 1.05 Set I-Source level (amps)
-210 to 210 Set V-Source level (volts)
(-21 to 21, 2400-LV and 2401)
DEFault 0A or 0V
MINimum -1.05A or -210V (-21V, 2400-LV
and 2401)
MAXimum +1.05A or +210V (+21V, 2400-LV
and 2401)
2410
<n> = -1.05 to 1.05 Set I-Source level (amps)
-1100 to 1100 Set V-Source level (volts)
DEFault 0A or 0V
MINimum -1.05A or -1100V
MAXimum +1.05A or +1100V
2420
<n> = -3.15 to 3.15 Set I-Source level (amps)
-63 to 63 Set V-Source level (volts)
DEFault 0A or 0V
MINimum -3.15A or -63V
MAXimum +3.15A or +63V
2425 and 2430 DC Mode
<n> = -3.15 to 3.15 Set I-Source level (amps)
-105 to 105 Set V-Source level (volts)
DEFault 0A or 0V
MINimum -3.15A or -105V
MAXimum +3.15A or +105V
2430 Pulse Mode
<n> = -10.5 to 10.5 Set I-Source level (amps)
-105 to 105 Set V-Source level (volts)
DEFault 0A or 0V
MINimum -10.5A or -105V
MAXimum +10.5A or +105V