18-78 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
Query :CURRent? Query programmed amplitude of I-Source
:CURRent? DEFault Query *RST default amplitude
:CURRent? MINimum Query lowest allowable amplitude
:CURRent? MAXimum Query highest allowable amplitude
:VOLTage? Query programmed amplitude of V-Source
:VOLTage? DEFault Query *RST default amplitude
:VOLTage? MINimum Query lowest allowable amplitude
:VOLTage? MAXimum Query highest allowable amplitude
Description This command is used to immediately update the amplitude of a
fixed source. This command is not valid for the list or sweep mode.
NOTE The sourcing :MODE command is used to select a fixed source. (See “Select
sourcing mode,” page 18-73.)
If a manual source range is presently selected, then the specified
amplitude cannot exceed that range. For example, if the V-Source is
on the 2V range (auto range disabled), you will not be able to set the
V-Source amplitude to 3V. In auto range, the amplitude can be set to
any level that is within the capabilities of the source.
The MINimum and MAXimum parameters are only valid if the highest
source range is presently selected. Sending the MINimum or MAXi
-
mum parameters on a lower source range will generate error -221
(Setting Conflict).
TRIGgered[:AMPLitude] <n>
:SOURce[1]:CURRent[:LEVel]:TRIGgered[:AMPLitude] <n> Set fixed I-Source amplitude when
triggered
:SOURce[1]:VOLTage[:LEVel]:TRIGgered[:AMPLitude] <n> Set fixed V-Source amplitude when
triggered
Parameters 2400/2400-LV/2401
<n> = -1.05 to 1.05 Set I-Source amplitude (amps)
-210 to 210 Set V-Source amplitude (volts)
(-21 to 21, 2400-LV and 2401)
DEFault 0A or 0V
MINimum -1.05A or -210V
MAXimum +1.05A or +210V
2410
<n> = -1.05 to 1.05 Set I-Source amplitude (amps)
-1100 to 1100 Set V-Source amplitude (volts)
DEFault 0A or 0V
MINimum -1.05A or -1100V
MAXimum +1.05A or +1100V
2420
<n> = -3.15 to 3.15 Set I-Source amplitude (amps)
-63 to 63 Set V-Source amplitude (volts)
DEFault 0A or 0V
MINimum -3.15A or -63V
MAXimum +3.15A or +63V