User Manual
2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-71
[:STATe] <b>
[:SENSe[1]]:AVERage[:STATe] <b> Enable or disable filter
Parameters <b> = 0 or OFF Disable digital filter
1 or ON Enable digital filter
Query :STATe? Query state of digital filter
Description These commands are used to enable or disable the digital averaging
filter. When enabled, voltage, current, and resistance readings are fil
-
tered according to how the filter is configured.
NOTE For the Model 2430 Pulse Mode, this command is invalid (error +831).
SOURce subsystem
This subsystem is used to configure and control the I-Source and V-Source, and to set
the logic level (high or low) of each digital output line. The commands for this sub-
system are summarized in Table 18-7.
SOURce[1]
Use the following commands to configure and control the I-Source and V-Source. At
the end of this subsystem are program examples of sweeps and lists.
Control source output-off
[:IMMediate]
:SOURce[1]:CLEar[:IMMediate] Turn source output off
Description This command is used to turn off the source output. The output will
turn off after all programmed source-measure operations are com
-
pleted and the instrument returns to the idle state.
Note that if auto output-off is enabled, the source output will automat-
ically turn off. (See next command.)