17-2 SCPI Signal Oriented Measurement Commands 2400 Series SourceMeter
®
User’s Manual
Command summary
The signal oriented measurement commands are used to acquire readings. You
can use these high-level instructions to control the measurement process. These
commands are summarized in Table 17-1.
Configuring measurement function
NOTE For the Model 2430, the :CONFigure commands are not valid while in
the Pulse Mode. They result in error +831: Invalid in Pulse Mode.
:CONFigure:<function>
Parameters <function> = CURRent[:DC] Amps function
VOLTage[:DC] Volts function
RESistance Ohms function
Query :CONFigure? Returns active function(s).
Description This command configures the instrument to a specific setup for
measurements on the specified function. The :READ? command
is then typically used to trigger a specified number of measure
-
ments. See “:READ?.”
When this command is sent, the SourceMeter will be configured
as follows:
Select specified function.
All controls related to the selected function are defaulted to
the *RST values.
The event control source of the Trigger Model is set to Imme-
diate.
Table 17-1
Signal oriented measurement command summary
Command Description
:CONFigure:<function>
:CONFigure?
:FETCh?
:READ?
:MEASure[:<function>]?
Configures SourceMeter for measurements on specified
function. Output turns on.
Returns active function(s).
Requests latest readings.
Performs an :INITiate and a :FETCh?.
One-shot measurement mode. Performs a :CONFigure and
a :READ?.