– SRC MEM LOC — Use this option with a Source Memory Sweep to
select the next memory location point in the sweep when the PASS
condition occurs. If NEXT is selected, the next point in the sweep list
will be selected. You can also branch to a different point in the sweep
by specifying the memory
LOCATION# (1 to 100).
• EOT MODE — Use this menu item to control the operation of Digital I/O
line 4 to act as an end-of-test (EOT) or BUSY signal:
– BUSY — Set Digital I/O line 4 HI while unit is busy. With BUSY
selected, the unit behaves as if it is in 3-bit mode. Requires SOT as the
arm source.
– /BUSY — Set Digital I/O line 4 LO while unit is busy. With /BUSY
selected, the unit behaves as if it is in 3-bit mode. Requires SOT as the
arm source.
– EOT — In 3-bit mode, automatically output a HI pulse on Digital I/O line
4 at end-of-test. In 4-bit mode, EOT is not automatically controlled.
Requires digital auto-clear to be enabled.
– /EOT — In 3-bit mode, automatically output a LO pulse on Digital I/O
line 4 at end-of-test. In 4-bit mode, EOT is not automatically controlled.
Requires digital auto-clear to be enabled.