2400 Series SourceMeter
®
User’s Manual Sweep Operation 10-17
Step 3: Set delay.
Set the source delay as follows:
1. Press CONFIG then SOURCE V or SOURCE I depending on the selected
source function.
2. Select DELAY, then press ENTER.
3. Set the delay to the desired value, then press ENTER.
4. Press EXIT to return to normal display.
Step 4: Turn output on.
Press the ON/OFF OUTPUT key to turn the output on (OUTPUT indicator turns
on). The SourceMeter will output the programmed bias level.
Step 5: Run sweep.
To run the sweep, press the SWEEP key. After the sweep is completed, turn the
output off by pressing the ON/OFF OUTPUT key.
Step 6: Read buffer.
Use the RECALL key to access the source-measure readings stored in the buffer.
Use the TOGGLE to display statistical information.
Performing a source memory sweep
Step 1: Store setups in source memory.
Store instrument setups in source memory as follows:
1. Configure the SourceMeter for various desired operating modes such as
source, measure, delay, and/or math expression operation. See
Table 10-2
for settings that can be stored in each source memory location.
2. Press MENU to display the MAIN MENU:
Select SAVESETUP.
Select SOURCE MEMORY.
Select SAVE.
•Use the ¹ and ƒ keys, and the cursor keys to display the desired mem-
ory location, and press ENTER.
Use the EXIT key to back out of the menu structure.
3. Repeat Steps 1 and 2 for all points in the sweep.