User Manual

7-10 Range, Digits, Speed, and Filters 2400 Series SourceMeter
®
User’s Manual
Front panel filter control
Configuring filter
Filter type and count is configured from the CONFIGURE FILTERING menu and
is structured as shown in Figure 7-2. Use Section 1, “Rules to navigate menus,” to
check and/or change the filter configuration.
CONFIGURE FILTERING menu
Press CONFIG and then FILTER to display the menu.
AVERAGING MODE — Use this menu item to select filter type
(MOVING or REPEAT).
AVERAGE COUNT — Use this menu item to specify filter count
(1 to 100 readings).
NOTE The configured filter is the same for all measurement functions.
Enabling filter
The filter is enabled by pressing the FILTER key. The “Filter Enabled” message
and the filter count are briefly displayed. The FILT annunciator is on while the filter
is enabled. Pressing FILTER a second time disables filter.