User Manual

7-6 Range, Digits, Speed, and Filters 2400 Series SourceMeter
®
User’s Manual
Range and digits programming example
Table 7-2 shows a programming example for controlling range and digits. The
SourceMeter is set up as follows:
Source function: volts
Source level: 10V
Measure function: amps
Amps range: 10µA
Display digits: 5H
Table 7-2
Range and digits programming example
Command Description
*RST
:SOUR:FUNC VOLT
:SOUR:VOLT 10
:SENS:FUNC “CURR”
:SENS:CURR:RANG 10E-6
:DISP:DIG 5
:OUTP ON
:READ?
:OUTP OFF
Restore GPIB defaults.
Volts source function.
Output 10V.
Amps measure function.
10µA range.
5H display digits.
Turn on output.
Trigger and acquire reading.
Turn off output.