User’s Manual Range, Digits, Speed, and Filters 7-5
Digits
The display resolution of the measured reading depends on the DIGITS setting.
This setting is global, which means the digits setting selects display resolution for
all measurement functions.
The DIGITS setting has no effect on the remote reading format. The number of
displayed digits does not affect accuracy or speed. Those parameters are con-
trolled by the SPEED setting.
Setting display resolution
There are two ways to set display resolution:
DIGITSPress the DIGITS key until the desired number of digits is dis-
played.
CONFIG DIGITSPress CONFIG and then DIGITS to display the digits
menu. Place the cursor on the desired number of digits (3.5, 4.5, 5.5, or
6.5) and press ENTER.
NOTE The concurrent measurement (available on the secondary display by
using the TOGGLE key) is always 5
H
digits.
Changing SPEED changes DIGITS, but changing DIGITS does not
change SPEED.
Remote range and digits programming
Table 7-1 summarizes the commands necessary to control range and digits. See
Section 18 for more details on these commands.
Range and digits commands
Table 7-1
Range and digits commands
Command Description
:SENSe:CURRent:RANGe <n>
:SENSe:CURRent:RANGe:AUTO <state>
:SENSe:VOLTage:RANGe <n>
:SENSe:VOLTage:RANGe:AUTO <state>
:SENSe:RESistance:RANGe <n>
:SENSe:RESistance:RANGe:AUTO <state>
:DISPlay:DIGits <n>
Select manual amps range (n = range).
Enable/disable auto amps range (state = ON or OFF).
Select manual volts measure range (n = range).
Enable/disable auto volts range (state = ON or OFF).
Select manual ohms range (n = range).
Enable/disable auto ohms range (state = ON or OFF).
Set display digits (n = 4, 5, 6, or 7).
Note: For the Model 2430, the :AUTO range commands are not valid while in the Pulse Mode.