User Manual

2400 Series SourceMeter
®
User’s Manual Source-Measure Concepts 6-25
Figure 6-15
Source V
Measure only (V or I)
Figure 6-16 shows the configurations for using the SourceMeter exclusively as a
voltmeter or ammeter. As shown in Figure 6-16A, the SourceMeter is configured
to measure voltage only by setting it to source 0A and measure voltage.
CAUTION V-Compliance must be set to a level that is higher than the mea-
sured voltage. Otherwise, excessive current will flow into the
SourceMeter. This current could damage the SourceMeter. Also,
when connecting an external voltage to the I-Source, set the output-
off state to the high-impedance mode. (See
Section 13, “Output-off
states.”)
In Figure 6-16B, the SourceMeter is configured to measure current-only by setting
it to source 0V and measure current. Note that in order to obtain positive (+) read-
ings, conventional current must flow from IN/OUT HI to LO.
Remote
SENSE HI
IN/OUT HI
Remote
I-Meter
IN/OUT LO
V-Meter
Sense Output
Adjust V-Source
(Feedback)
SENSE LO
Local
Local
x1
GUARD SENSE
GUARD
+
V-Source